Compiled Documents for Wro 300 Operators Manual. Updated Title Size 31 Mar 2015: Gambro 300 Wro Operation Manual.pdf Full Version: kx80 owners manual WRO 300, WRO 300 H units. Efficient water purification at your fingertips. Our complete understanding of water purification in a powerful single-patient unit. Instructions wro300andwro300hinstructionsrev02-2006.pdf 46 pages (1MB). GAMBRO GAMBRO Water Purification Units WRO 300 and WRO 300 H Instructions Rev 02.

WRO 300 Technical Updates

Gambro Wro 300 Manual

Gambro Wro 300 Manual

WRO 300 General Information

Physical Dimensions
Height:
563 mm
Width: 205 mm
Base:
185 mm x 380 mm
Weight:29 kg

Water Supply

Input:Min. 3.0 l/min required

Feed Water Pressure:22 psi to 44 psi (0.15 to 0.3 MPa during operation.)

Feed Water Temp: +5 to +30 °C

Power Supply

Main Voltage: 100-240 VAC (line voltage range 85 to 264 VAC)

Frequency: 50/60 Hz (line frequency range 45 to 65 Hz)

Power:570 W

Conductivity Measurement

Product Water Conductivity: 1 - 500 uS/cm at ±10% or ±10 uS/cm which ever is greater.

Feed Water Conductivity: 1 - 2000 uS/cm at ±10% or ±10 uS/cm which ever is greater.

Product Water

Output Flow: Min. 1.1 l/min at +10°C

Product Water Loop:20 meters (2 x 10 meters) Product Water Pressure: 0.12 to 0.6 MPa during RUN mode.

Quality:Depends on feed water quality. If potable water is used, the following rejection rates will be obtained:
Total dissolved salts: > 96%
Bacteria and endotoxin: > 99%

Drain

Drain Output:1.2 ±0.1 l/min

Drain Height:Max. 1.0 m

Drain Hose Length:Max. 5 m

Drain Flow: Min. 3.0 l/min required

Gambro Wro 300 Manual Pdf

Included Features

• Automated processes ensure consistency with disinfection procedures.

• Programmable automated flushing and continuous product water loop eliminate stagnation and ensures system cleanliness.

• Continuous product water loop minimizes water volume requirements for purified water production.

• Portable, compact and easy to operate with a very low noise level.

• Additional automated procedures and internal diagnostic minimize home patients and caregiver involvement.

• End-to-end disinfection with dialysis machines.